Logo Caen

AG523

6 Channel 12 kV/1 mA Common Ground Board

Features

  • 6 independently controllable High Voltage channels
  • 12 kV / 1 mA output range
  • Available with either positive or negative polarity
  • CPE HV coaxial connectors
  • Low ripple
  • Under/over-voltage alert, overcurrent and max. voltage protection
  • Interlock logic for unit enable
  • Software Tool for easy channel management
3 Years Warranty
placeholder
placeholder

Overview

The CAEN Mod. AG523 is a double width board (10 TE wide) that houses 6 Independent high voltage channels available with either positive or negative polarity. The channels share a Common Ground.

APP PC5933T 6 CPE HV coaxial connector

Double width (10 TE wide), 6 channel for Mod. AG523

Consult our connectors reference page for technical information.

The output voltage can be programmed and monitored in the 0 ÷ 12 kV range with 1 V monitor resolution.

The Maximum current full scale is 100 μA, with 10 nA Iset/Imon monitor resolution.

The board features the CPE HV output connectors and the relevant coaxial cable connectors (Mod. A997) are also available.

Independently programmable for each channel:

Output voltage: 0 ÷ 12 kV step: 1 V
Current limit 0 ÷ 100 μA step: 10 nA
V Ramp up/down: 1 ÷ 500 V/sec step: 1 V/sec
TRIP parameter 0 ÷ 999.9 s; 1000 s = Infinite step: 0.1 s

Safety features allows the module to perform as a current generator and includes:

Channels
can be enabled or disabled through the Global Interlock logic.
Overvoltage and Undervoltage warning when the output voltage differs from the programmed value.
Overcurrent detection when a channel attempts to exceed the programmed current limit, it signalled to be in “overcurrent” and enter in a TRIP status. The output voltage is varied to keep the current below the programmed limit for a programmable TRIP time, then the channel is switched off. If TRIP is set to “constant current mode”, the channel behaves like a current generator.
Hardware VMAX
maximum output voltage can be set, via front panel potentiometer, at the same common value for all the board channels. VMAX value can be read out via software.
Safety Board Interlock this protection disables the HV generation when the HV outputs are not connected to their loads.

Accessories available:

  • HV coaxial cable connector (Mod. A997)

Technical Specifications

No. of Channels

6 (Common Ground)

Output Voltage

0÷12 kV

Polarity

Positive / Negative depending on purchased version

Max. Output Current

1 mA

Voltage Set Resolution

1 V

Voltage Monitor Resolution

1 V

Current Set Resolution

100 nA

Current Monitor Resolution

100 nA

VMAX hardware

0÷12 kV common for all the board channels

VMAX hardware accuracy

± 2% of FSR

VMAX software

0÷12 kV settable for each channel

VMAX software resolution

1 V

Ramp Up/Down

1÷500 Volt/sec, 1 Volt/sec step

Voltage Ripple

< 30 mVpp

Voltage Monitor vs. Output Voltage Accuracy

± 0.3% ± 2 V

Voltage Set vs. Voltage Monitor Accuracy

± 0.3% ± 1 V

Current Monitor vs. Output Current Accuracy

± 2% ± 1 µA

Current Set vs. Current Monitor Accuracy

± 2% ± 0.1 µA

Compare

Compare with High Voltage Power Supply.

Loading...

Accessories

A997
HV coaxial cable connector for A1526
HV Cables
High Voltage Cable Assemblies

Ordering Options

Code Description
WAG523XAAAAA AG523N - SYx527 H.V. channels -12 KV 1 mA - CPE Conn. common ground (6 ch 10TE wide)   RoHS
WAG523XPAAAA AG523P - SYx527 H.V. channels +12 KV 1 mA - CPE Conn. common ground (6 ch 10TE wide)   RoHS